Features:
- DC measurement accuracy of 0.0024% for outstanding reliability.
- Industry-leading thermocouple accuracy of 0.5 °C.
- Supports up to 66 universal differential isolated inputs.
- On-screen color trend graphing for simultaneous viewing of up to four channels.
- User-friendly menu system simplifies setup and data management.
- Comprehensive input types including: ac V, dc V, ac I, dc I, thermocouple, PRT (2,3,4 w), thermistor, resistance (2-4 w), frequency.
- Multi-channel real-time data display enhances monitoring capabilities.
- Six-and-a-half digit bench DMM function available for front-panel inputs.
- Monitor function allows for real-time viewing and charting between scans.
- Includes 20 on-board math channels for added analytical capacity.
- One-button screen capture function for easy data preservation.
- Impressive 45 channels/second basic DC scan rate.
- Robust internal memory with 57,000 scan/setup files.
- Supports USB flash drives for additional data storage options.
- Inbuilt data security features for safeguarding sensitive information.
- Meets a maximum input safety rating of 300 V CAT II.
Specifications:
- Technical Specifications:
- DC Volts:
- Resolution: 0.1 µV to 1 mV
- AC Volts:
- Resolution: 0.1 µV to 1 mV
- Resistance:
- Resolution: 0.1 mΩ to 100 Ω
- Frequency:
- RTD (Pt 100):
- Thermocouples (J, K, T and Others):
- J: Range -200 °C to 1,000 °C, Resolution: 0.01 °C, Accuracy: 0.53 °C
- K: Range -200 °C to 1,000 °C, Resolution: 0.01 °C, Accuracy: 0.60 °C
- T: Range not specified, Resolution: 0.01 °C, Accuracy: 0.62 °C
- Other types (R, S, N, E, B, C, D, G, L, M, U, W) also supported.
- Channel Capacity: 22 to 66 analog inputs and 8 digital I/O.
- 6 TTL alarm outputs, configurable to one or multiple channels.
- Power: 100V to 264V in multiple power selections (100V, 120V, 220V, 240V) with a frequency of 47Hz to 440Hz.
- Power consumption: 34VA peak, 24W average.
- Communication: LAN support (10/100 Mb TCPIP DHCP) with SCPI command set for remote commands.
- Memory: Up to 57,000 scans and 1,000 setup files.
- File Security: Dual-level security with an Administrator and five secure user profiles.
- Configurable alarm setpoints (2 per channel) for high/low alerts.