Fluke 754 Documenting Process Calibrator

Do you have a Fluke 754 Documenting Process Calibrator you want to sell?

We buy used equipment! Learn about selling

Image Coming Soon

Fluke 754 Documenting Process Calibrator

The Fluke 754 Documenting Process Calibrator is a robust and versatile handheld device that consolidates multiple tools into one compact unit. It excels in sourcing, simulating, and measuring various parameters including pressure, temperature, and electrical signals, making it an indispensable asset for calibration professionals. This multifunction calibrator offers the capability to download and upload procedures, instructions, and data for seamless printing, archiving, and analysis. The integrated HART® interface facilitates efficient calibration, maintenance, and troubleshooting of HART and other instrumentation.

Key Features:

    • Measures volts, mA, RTDs, thermocouples, frequency, and ohms for comprehensive sensor testing.
    • Sources and simulates volts, mA, thermocouples, RTDs, frequency, ohms, and pressure for precise transmitter calibration.
    • Powers transmitters during testing with loop supply while simultaneously measuring mA.
    • Utilizes any of the 29 Fluke 700Pxx Pressure Modules for pressure measurement or sourcing.
    • Configures automated as-found/as-left procedures to comply with quality programs or regulatory standards.
    • Records and accurately documents calibration results, retaining data for up to one week.
    • Features include autostep, custom units, user-defined values, one-point and two-point switch testing, and more.
    • User-friendly design with a bright white dual display for simultaneous parameter viewing.
    • Supports a multi-lingual interface, enhancing accessibility for global users.
    • Equipped with a rechargeable Li-Ion battery, providing 10 hours of uninterrupted operation.
    • Handles fast pulsed RTD transmitters and PLCs with pulse durations as short as 1 ms.
    • Incorporates DPC/Track Sample software for enhanced functionality.
    • Compatibility with various Asset Management software packages streamlines integration.
    • Eliminates the need for external boxes or additional tools for routine HART calibration.
    • Offers fast HART communication and supports popular HART transmitter models.
    • Allows interrogation to determine device type, manufacturer, model, and tag.
    • Facilitates sensor mapping reconfiguration for dual sensor temperature transmitters.
    • Reads and writes HART configuration functions for efficient field adjustments.

Specifications:

Measurement Accuracy
Voltage DC Range/Resolution 1 Year 2 Years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Current DC 30.000 mA 0.01% + 5 µA 0.015% + 7 µA
110.00 mA 0.01% + 20 µA 0.015% + 30 µA
Resistance 10.000 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
100.00 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
1.0000 kΩ 0.05% + 500 mΩ 0.07% + 0.5 Ω
10.000 kΩ 0.1% + 10 Ω 0.15% + 15 Ω
Frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz 0.05 kHz
Source Accuracy
1 Year 2 Years
Voltage DC 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Current DC 22.000 mA (source) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current sink (simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance 10.000 Ω 0.01% + 10 mΩ 0.015% + 15 mΩ
100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ
1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω
Frequency 0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure
Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length 8000 readings (7980 for 30 or 60 readings/minute)
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Rate 4 steps/second
Trip detect Continuity or voltage (continuity detection not available when sourcing current)
Loop power function Voltage Selectable, 26 V
Accuracy 10%, 18 V minimum at 22 mA
Maximum current 25 mA, short circuit protected
Maximum input voltage 50 V DC
Step functions Source functions Voltage, current, resistance, frequency, temperature
Manual step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, step value, time per step, repeat
Environmental Specifications
Operating temperature -10°C to +50°C
Storage temperature -20°C to +60°C
Dust/water resistance Meets IP52, IEC 529
Operating altitude 3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight 1.2 kg (2.7 lb)
Batteries Internal Battery Pack Li-ion: 7.2V, 4400mAh, 30 Wh
Battery life >8 hours typical
Battery replacement Replace without opening calibrator; no tools required
Side port connections Pressure module connector
USB Connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Data storage capacity 1 week of calibration procedures results