Keysight / Agilent E5270A Parametric Measurement System Mainframe, 8-Slot

Do you have a Keysight / Agilent E5270A Parametric Measurement System Mainframe, 8-Slot you want to sell?

We buy used equipment! Learn about selling

Image Coming Soon

Keysight / Agilent E5270A Parametric Measurement System

The Agilent E5270A is an advanced 8-slot Parametric Measurement Mainframe designed for flexibility and precision, making it an essential tool for current-voltage characterization of diverse materials and devices. This user-configurable system allows the installation of up to eight single-slot modules (such as the MPSMU) or up to four dual-slot modules (like the HPSMU), catering to a range of testing requirements.

Key Features:

    • High-speed DC parametric measurements for enhanced efficiency
    • Configurable with eight slots for plug-in modules to suit varying measurement needs
    • User-friendly front panel interface for quick spot measurements
    • Equipped with a high-resolution, analog-to-digital converter (ADC), accessible by all installed modules
    • Each SMU module includes a high-speed ADC, enabling superior measurement accuracy
    • 4-Amp ground unit for stable operation
    • BNC trigger-in and trigger-out connectors for seamless integration
    • 16 general-purpose digital I/Os to enhance connectivity options
    • Programmable memory to store complex measurement sequences
    • GPIB port supports instrument control for automated testing
    • Features self-test, self-calibration, and diagnostics for reliable operation

Supported Measurement Modes:

    • Spot Measurements
    • Pulsed Spot Measurements
    • Quasi-Pulsed Spot Measurements
    • Staircase Sweep Measurements
    • Multi-Channel Sweep Measurements
    • Pulsed Sweep Measurements
    • Staircase Sweep with Pulsed Bias
    • Linear Search Measurements
    • Binary Search Measurements

Specifications:

    • Measurement Range:
      • With E5280A: 2 mV to 200 V, 10 fA to 1 A
      • With E5281A: 2 mV to 100 V, 10 fA to 200 mA
    • Maximum Output Power: 80W
    • Maximum Common to Ground Voltage: ≤ 42 V
  • Pulse Measurement:
      • Pulse Width: 500 µs to 2 s
      • Pulse Period: 5 ms to 5 s
      • Period ≥ Width + 2 ms (when Width ≤ 100 ms)
      • Period ≥ Width + 10 ms (when Width > 100 ms)
      • Pulse Resolution: 100 µs
    • GNDU: Ground Unit included with the E5270A mainframe
    • Output Voltage: 0 V ±100 mV
    • Maximum Sink Current: 4 A
    • Output Terminal/Connection: Triaxial connector, Kelvin (remote sensing)