Keysight / Agilent E5270B Parametric Measurement Mainframe
The Agilent E5270B is an advanced 8-slot parametric measurement mainframe designed for complete user configurability, allowing the installation of up to eight single-slot modules (such as the E5281B MPSMU) or up to four dual-slot modules (such as the E5280B HPSMU), catering to diverse measurement requirements.
Precision Current-Voltage Characterization
Leveraging the best in class current measurement performance, with sensitivity as low as 1 nA, the Keysight E5270B provides an all-in-one solution for comprehensive current-voltage characterization across a wide array of materials and electronic devices. Its modular architecture facilitates easy configuration or upgrades, ensuring the system meets evolving experimental needs.
Included with the device is the EasyEXPERT group+ GUI-based characterization software, which streamlines all measurement tasks from setup through to detailed data analysis, delivering unparalleled reliability and efficiency for device evaluation, including semiconductors and passive components.
Key Features
- High-speed, dc parametric measurements
- Configurable with eight slots for various plug-in modules
- User-friendly interface for quick spot measurements via the front panel
- High-resolution, analog-to-digital converter (ADC) available across all installed modules
- Integrated high-speed ADC on each installed SMU
- 4-Amp Ground Unit integrated
- BNC trigger-in and trigger-out connectors
- 16 general-purpose digital I/O ports
- Ample program memory for diverse applications
- GPIB port for seamless instrument control
- Self-test, self-calibration, and diagnostic capabilities
Current-Voltage Measurement Capabilities
- Accurate measurement ranges from 0.1 fA to 1 A and 0.5 µV to 200 V
- Spot and sweep measurement capabilities included
- Pulse spot and sweep measurements with a minimum pulse width of 500 µs
- Compatible with ASU (atto-sense and switch unit) for enhanced measurement resolution
Supported Measurement Modes
- Staircase Sweep with Pulsed Bias
Specifications
- Measurement range:
- With E5280B: -200 V to 200 V, -1 A to 1 A (resolution: 2 µV, 10 fA)
- With E5281A: -100 V to 100 V, -100 mA to 100 mA (resolution: 0.5 µV, 10 fA)
- Maximum Output Power: 80 W
- Maximum Common to Ground Voltage: ≤ 42 V
- Pulse Measurement:
- Pulse width: 500 µs to 2 s
- Pulse period: 5 ms to 5 s
- Period ≥ Width + 2 ms (Width ≤ 100 ms)
- Period ≥ Width + 10 ms (Width > 100 ms)
- Ground Unit: Furnished with E5270B mainframe
- Output Voltage: 0 V ± 100 mV
- Maximum Sink Current: 4 A
- Output Terminal/Connection: Triaxial connector, Kelvin (remote sensing)