The Keysight N6784A is a cutting-edge source/measure unit (SMU) tailored for general-purpose precision sourcing and measurement applications. This versatile instrument is expertly engineered to meet the demands of a broad spectrum of industries and devices under test (DUTs), making it an essential addition to any test environment. The N6784A ensures glitch-free operation, allowing for safe adjustments during output and measurement range changes—even with capacitive loads up to 150 μF.
Featuring 4-quadrant operation, the N6784A can function as both a bipolar power supply and a bipolar electronic load, giving engineers unmatched flexibility in their testing procedures. With this SMU, you can be assured of reliable performance that enhances both test systems and lab benches.
Key Specifications
Output Ratings:
Programming Accuracy (at 23°C ± 5°C after 30-minute warm-up):
- Voltage, 20 V range: 0.025% + 1.8 mV
- Voltage, 6 V range: 0.025% + 600 μV
- Current, 3 A and 1 A ranges: 0.04% + 300 μA
- Current, 10 mA range: 0.25% + 5 μA
Output Programming Speed:
- Output voltage can change in 10 μs or operate at up to 100 kHz into a resistive load.
Measurement Accuracy (at 23°C ± 5°C):
- Voltage, 20 V range: 0.025% + 1.2 mV
- Current, 3 A range: 0.03% + 250 μA
- Current, 100 mA range: 0.025% + 10 μA
- Current, 10 μA range: 0.025% + 8 nA
Features and Benefits:
- Glitch-free sourcing and measurement allow for output and measurement range changes without disruption.
- 4-quadrant operation provides functionality as a bipolar voltage/current source or as a bipolar CV/CC electronic load.
- Four current programming ranges for precise current sourcing down to microamperes.
- Four current measurement ranges facilitate accurate current capture across various DUT operating modes.
- Excellent transient response ensures voltage or current stability during rapid load changes.
- Fast modulation of DC enables waveform creation up to 100 kHz for DUT stimulation or loading.
- High-speed digitized measurements allow for power consumption monitoring of the DUT up to every 5 μs with a built-in 200 kHz digitizer.