Ohmeda Care Plus Infant Incubator

Do you have a Ohmeda Care Plus Infant Incubator you want to sell?

We buy used equipment! Learn about selling

Manufactured by: Datex-Ohmeda

Category: Incubators - Infant

Datex-Ohmeda Ohmeda Care Plus Infant Incubator
Datex-Ohmeda Ohmeda Care Plus Infant Incubator
Datex-Ohmeda None - Incubators - Infant
Datex-Ohmeda None - Incubators - Infant
Ohmeda Care Plus Infant Incubator
  • Advanced Neonatal Care Solution

The Ohmeda Care Plus Infant Incubator is designed to provide optimal care for neonates, ensuring a safe and stable environment crucial for their development. Although this particular model is currently unavailable, alternative options are available for hospital use, as well as specialized units for educational and simulation purposes.
Advanced Thermoregulation Technology
This incubator features a sophisticated microprocessor-based control system, enabling accurate and efficient thermoregulation. With the ability to maintain a precise temperature, caregivers can ensure that infants receive the stable conditions they need for optimal growth.
Key Features Include:
  • Microprocessor-Based Control System: Facilitates precise monitoring and adjustment of temperature settings for enhanced care.
  • Self-Test Functions: Automatically conducts tests at power-up and during normal operations to ensure optimal functionality.
  • Flexible Control Modes: Offers both patient control (Servo) mode and air control (Manual) mode to adapt to various clinical needs.
  • Adjustable Positioning: Continuous positioning capability from 0° to 12° allows for easy access and comfort for the infant.
  • Optional Active Double-Wall Construction: Provides an enhanced environment that further supports thermoregulation and patient stability.

Experience enhanced neonatal care with the Ohmeda Care Plus Infant Incubator, designed for the dedicated care of your tiniest patients. Explore our range of medical equipment and discover solutions that meet the demands of modern healthcare.

Ohmeda Care Plus Infant Incubator Specifications


Dimensions

  • Incubator with Elevating Base
  • Height: 54 - 62 in (136.2 - 157.5 cm)
  • Elevating Base Height: 35 - 43 in (88.9 - 109.2 cm) mattress to floor
  • Depth: 28 in (71.1 cm)
  • Width: 33 in (83.8 cm)
  • Weight: 195 lbs (89 kg)

  • Incubator with Stationary Pedestal
  • Height: 53 in (135.9 cm)
  • Depth: 28 in (71.1 cm)
  • Width: 33 in (83.8 cm)
  • Weight: 180 lbs (80 kg)

  • Drawer Packages for Elevating Base or Stationary Pedestal
  • Overall Dimensions: 11 x 11.5 x 18.25 in (28 x 29 x 46 cm)
  • Single Deep Drawer: 10 x 9 x 16.5 in (25 x 23 x 42 cm)
  • Dual Drawer: 4.5 x 9 x 16.5 in (11 x 23 x 42 cm)

  • Incubator with Cabinet
  • Height: 55 in (141 cm)
  • Depth: 25.5 in (64.6 cm)
  • Width: 35 in (88.9 cm)
  • Weight: 185 lbs (84 kg)

  • Incubator Characteristics
  • Casters: 5 in (12.7 cm) diameter, 2 locking, 2 non-locking
  • Mattress: 13.7 x 25.6 in (34.8 x 65 cm)
  • Mattress to Hood: 16 in (41 cm)
  • Door Height: 11 in (28 cm)
  • Positions: Continuous from 0° to 12°
  • Port Holes: 6
  • Tubing Ports: 6

Electrical

  • Voltage: 120 VAC, 60 Hz
  • Models: 115 V ± 10%, 5.7 amps
  • Compliance: Conforms to IEC 601.1, UL 544, and CSA 22.2 No. 125
  • Nominal Power Consumption: 450 watts at maximum heater output
  • Line Voltage Compensation: Monitors input voltage and adjusts heat output accordingly
  • Leakage Currents: Less than 100 microamps on 120V units
  • Leakage at Patient Probe: Less than 50 microamps on 120V units

Environment

  • Operating Temperature Range: 68° - 86°F (20° - 30°C)
  • Storage Temperature Range: 13° - 140°F (-25° - 60°C)
  • Operating Humidity Range: 0% to 90%

System Controls

  • Control System: Microprocessor-based, includes self-test functions at power-up and during normal operation
  • Patient Control (Servo) Mode: 95° - 98.6°F (35° - 37°C)
  • Air Control (Manual) Mode: 73.4° - 98.6°F (20° - 37°C), with capacity up to 102.2°F (39°C) using control panel override switch
  • Air Velocity: Average < 10 cm/sec with inner wall
  • Temperature Variability: Less than ± 0.2°F (0.1°C)
  • Resolution: ± 0°F (0.1°C)
  • Probe or Interchangeability: ± 0.2°F (0.1°C)
  • Average Oxygen Input Concentration Range:
  • 5 L/min: 25-45%
  • 10 L/min: 35-65%
  • 15 L/min: 45-70%
  • Humidification:
  • Standard: 10% to 75% dependent on nursery environment and incubator temperature setting
  • Servo: 40% to 85% regardless of nursery environment and incubator temperature setting