AC Adapter |
Universal 50-60 Hz, 100-240 VAC (Cat. No. STARA-PWR) |
Accuracy (Temperature) |
±0.1°C |
Accuracy (DO - Polarographic) |
±0.2 mg/L or ±2% of reading |
Accuracy (Dissolved Oxygen - RDO) |
±0.1 mg/L up to 8 mg/L; ±0.2 mg/L from 8 to 20 mg/L; ±10% of reading up to 50 mg/L |
Barometric Pressure Correction |
400 to 850 mmHg, automatic correction with built-in barometer or manual entry |
Measurement Modes |
Dissolved oxygen as % saturation or mg/L, with temperature and barometric pressure |
Display Type |
Color graphic LCD with customizable user options |
Ranges |
DO: 0.00 to 90.00 mg/L; RDO: 0.00 to 50.00 mg/L; Temperature: 0.0 to 50.0°C |
Dimension (L x W x H) |
28 cm x 24 cm x 10 cm (11" x 9.5" x 3.9") |
Weight |
1.2 kg (without modules), additional 0.2 kg per module |