Yokogawa 701251 Analog Voltage Input Module
The Yokogawa 701251 Analog Voltage Input Module is engineered to elevate your measurement capabilities with its dual-channel design that ensures high speed and excellent isolation for reliable data acquisition. This state-of-the-art module is built for professionals and industries that require precise readings in various environmental conditions.
Key Features:
- High Speed: Achieving a remarkable sample rate of 1 Ms/s.
- High Sensitivity Range: Capable of detecting signals as low as 10 mV.
- Low Noise Level: Offers a typical residual noise level of ±100 μV.
- Superior Noise Immunity: Designed to minimize the impact of external interferences.
Specifications:
- Effective Measurement Range: Twice of the setting range for enhanced versatility.
- Input Channels: 2 channels for simultaneous measurements.
- Input Coupling: Configurable for AC, DC, and GND.
- Maximum Sample Rate: Up to 1 MS/s.
- Input Format: Isolated unbalanced connections.
- Frequency Characteristics:
- 50 mV to 200 V Range: DC to 300 kHz.
- 20 mV and 10 mV Range: DC to 200 kHz.
- Voltage Range Setting: 10 mV to 200 V with 1-2-5 steps.
- Maximum Input Voltage:
- With 700929 (10:1): 600 V (DC + ACpeak).
- With 701901 + 701954 (1:1): 140 V (DC + ACpeak).
- Direct input or cables not meeting safety standards: 140 V (DC + ACpeak).
- Common Mode Voltage:
- With 700929 (10:1) / 701901 + 701954 (1:1): 400 Vrms (CAT l), 300 Vrms (CAT ll).
- For inputs not adhering to safety standards: 42 V (DC + ACpeak) (CAT I and II).
- Vertical Axis Accuracy:
- 50 mV to 200 V Range: ±(0.25% of range).
- 20 mV Range: ±(0.3% of range).
- 10 mV Range: ±(0.5% of range).
- Input Connector: BNC connector (isolated type).
- Input Impedance: 1 MΩ ± 1%, approximately 35 pF.
- Common Mode Rejection Ratio: 80 dB (50/60 Hz) or more (typical).
- Withstand Voltage: 1500 Vrms for 1 minute (across each terminal and earth).
- Insulation Resistance: 500 VDC, 10 MΩ or more across each terminal and earth.
- A/D Conversion Resolution: 16-bit (2400 LSB/range).
- Temperature Coefficient:
- 50 mV to 200 V Range: ±(0.02% of range)/ºC (typical).
- 20 mV Range: ±(0.05% of range)/ºC (typical).
- 10 mV Range: ±(0.10% of range)/ºC (typical).
- 10 mV to 200 V Range: ±(0.02% of range)/ºC (typical).
- Bandwidth Limit: Options include OFF, 40 kHz, 4 kHz, and 400 Hz.
Cut-off Characteristics: -12 dB/OCT (typical).
- Probe Attenuation Settings:
- Voltage Probes: 1:1, 10:1, 100:1, 1000:1.
- Current Probes: 10 A:1 V and 100 A:1 V variations.
- Reference Junction Compensation Accuracy:
- K, E, J, T, L, U, N: ±1°C.
- Input Bias Current: 20 nA or less.