Stanford Research SR850 DSP Lock-In Amplifier, 1 mHz to 102.4 kHz

Do you have a Stanford Research SR850 DSP Lock-In Amplifier, 1 mHz to 102.4 kHz you want to sell?

We buy used equipment! Learn about selling

Image Coming Soon

RENT Stanford Research Systems SR850 DSP Lock-In Amplifier (call for availability)

Introducing the Stanford Research Systems SR850 DSP Lock-In Amplifier, an advanced tool designed with innovative Digital Signal Processing (DSP) architecture. This exceptional device offers notable performance benefits over traditional lock-ins, including enhanced dynamic reserve, minimized drift, reduced distortion, and superior phase resolution. The built-in CRT display and extensive 65,536 points of memory enable versatile data display and processing options that are simply unattainable with standard models.

    • Frequency Range: 1 mHz to 102.4 kHz
    • Dynamic Reserve: >100 dB
    • Phase Resolution: 0.001 degree
    • Time Constants: Adjustable from 10 µs to 30 ks (up to 24 dB/oct rolloff)
    • Auto Features: Auto-gain, -phase, -reserve, and -offset
    • Data Logging: Up to 65k samples with smoothing, curve fitting, and statistical analysis
    • Connectivity: GPIB, RS-232, and 3.5" disk drive for seamless data handling
  • Digital Precision:
      • 18-bit A/D Converter at input
      • High-Speed DSP Chip for digitally synthesizing the reference oscillator and implementing precise mathematical calculations
    • Digital Flexibility: 7" CRT display supporting a variety of display options
  • Input Channel:
      • Differential input with 6 nV/vHz input noise
      • Input impedance: 10 MO, minimum full-scale input voltage sensitivity: 2 nV
      • No need for tracking band-pass filter at input
  • Reference Channel:
      • Can utilize an externally applied sine or square wave, or an internal digitally synthesized reference
      • Internal source boasts minimal reference phase noise, available for fixed or swept operation
  • High Dynamic Reserve:
      • Unmatched 100 dB dynamic reserve without compromising stability, accuracy, or introducing analog band-pass filter noise
  • Traces and Displays:
      • Storage for up to four user-defined traces with real-time operations
      • Display data visually with bar graphs, strip charts, and polar plots
      • Autoscale feature for quick optimization of display settings
      • Configurable screen formats, including single large or dual display
  • Convenient Auto Measurements:
      • Quick access auto functions for gain, phase, dynamic reserve, and display scaling
      • Streamlined configuration process utilizing auto options
  • Auxiliary A/Ds and D/As:
      • Four rear-panel A/D inputs for external signals; values incorporated into SR850 trace definitions
      • Four D/A outputs available for fixed or synchronized voltage levels
  • Analysis Features:
      • Savitsky-Golay smoothing and statistical metrics for trace regions
      • Curve fitting capabilities for diverse data sets, including linear and Gaussian fits
      • Trace calculator enables arithmetic and trigonometric operations on trace data
  • Interfaces and Hardcopies:
      • Standard RS-232 and GPIB interfaces for comprehensive control and querying
      • Hardcopy options include screen dumps to printers and plots to HP-GL compatible devices
  • Signal Channel:
      • Voltage Inputs: Single-ended or differential, with sensitivities ranging from 2 nV to 1 V
      • Current Input: Options for 106 or 108 V/A
      • Input Impedance: 10 MO + 25 pF for voltage inputs; 1 kO for current
      • Noise Levels: 6 nV/vHz at 1 kHz; down to 0.013 pA/vHz at 100 Hz for current inputs
  • Reference Channel Specs:
      • Frequency Range: 0.001 Hz to 102.4 kHz with reference input capabilities
      • Exceptional phase accuracy and low noise levels
      • Phase drift kept below 0.01°/°C below 10 kHz
  • Demodulator Performance:
      • Highly stable outputs with digital outputs showing no drift
      • Offset adjustable by ±100 % and can expand up to 256×
  • Internal Oscillator:
      • Range from 1 mHz to 102.4 kHz with an impressive amplitude accuracy of 1 %
      • Phase-locked output with linear and logarithmic sweep capabilities
  • Inputs and Outputs:
      • X,Y Outputs: ±10 V, updated at 256 ksamples/s
      • Auxiliary A/D and D/A Inputs/Outputs with ±10 V range
      • Full suite of interface options: IEEE-488.2, RS-232, Centronics
  • Display Configurations:
      • Support for single and dual displays with real-time trace definitions
      • Comprehensive data buffer for storing up to 64k data points
  • General Specification:
      • Hardcopy and data storage capabilities with options for screen dumps
      • Equipped with a 3.5" MS-DOS compatible disk drive for data and setup management
      • Power requirements: 60 W, compatible with 100/120/220/240 VAC, 50/60 Hz

How can we help you with today?